大象传媒

Acid-alkali titrations

The concentration of an or can be calculated by carrying out an experiment called a .

Materials

The apparatus needed includes:

  • a pipette to accurately measure a certain of acid or alkali
  • a pipette filler to use the pipette safely
  • a conical flask to contain the liquid from the pipette
  • a burette to add small, measured volumes of one reactant to the other reactant in the conical flask
Apparatus needed for titration: burette, pipette, conical flask and pipette filler.
Figure caption,
Apparatus needed to carry out a titration

Method

This is an outline method for carrying out a titration in which an acid is added to alkali.

  1. Use the pipette and pipette filler to add 25 cm3 of alkali to a clean conical flask.
  2. Add a few drops of and put the conical flask on a white tile.
  3. Fill the burette with acid and note the starting volume.
  4. Slowly add the acid from the burette to the alkali in the conical flask, swirling to mix.
  5. Stop adding the acid when the end-point is reached (the appropriate colour change in the indicator happens). Note the final volume reading.

The same method works for adding an alkali to an acid 鈥 just swap around the liquids that go into the conical flask and burette.

The titre [Higher tier 鈥 GCSE Chemistry only]

The difference between the reading at the start and the final reading gives the volume of acid (or alkali) added. This volume is called the titre.

For example, if the reading at the start is 1.00 cm3 and the final reading is 26.50 cm3, then the titre is 25.50 cm3 (26.50 鈥 1.00). The titre will depend upon the volume of liquid in the conical flask, and the concentrations of the acid and alkali used. Repeat the titration until you have at least two readings that are within 0.20 cm3 of each other.

Titration results are recorded in a table like this:

TitrationRough123
End vol (cm3)26.8024.6024.5025.00
Start vol (cm3)1.000.000.400.20
Titre (cm3)25.8024.6024.1024.80
TitrationEnd vol (cm3)
Rough26.80
124.60
224.50
325.00
TitrationStart vol (cm3)
Rough1.00
10.00
20.40
30.20
TitrationTitre (cm3)
Rough25.80
124.60
224.10
324.80

You can ignore the first (rough) titration result. Titrations one and three have results that are within 0.20 cm3 of each other but the result for titration two is too low. You would calculate the mean titre for titrations one and three, and ignore titration two:

Mean titre = (24.60 + 24.80) 梅 2 = 49.40 梅 2 = 24.70 cm3

More guides on this topic